DF-110E Maintenance Free Oxygen Analyzer for OEMs




Flexible, Accurate and Reliable Oxygen Analysis to Meet Your OEM Requirements
  • Instantaneous response to oxygen change- less than 10 seconds to read 90% of a step change
  • Flexibility - put together the exact configuration you need to include in your system
  • Performance - the analyzer is delivered ready to use
  • Low/No Maintenance - no recalibration - no sensor replacement necessary
  • STAB-ELT option eliminates acid scrubbing for trace level measurements

The DF-110E offers a flexible and cost-effective way to include superior oxygen analysis in your OEM system. The DF-110E is calibrated to standards traceable to NIST and operated for several days to ensure flawless performance out-of-the-box. Depending on the application, the sensor can operate for years without recalibration and requires only the occasional addition of distilled water for maintenance.

DF-110E Specifications
  • Range
  • Ranges are available from 0-10 ppm to 0-25%

Sample Gas Compatibility
  • STANDARD SENSOR: All inert and passive gases including N2, CO, H2, Ar, freons, hydrocarbons, etc.
  • SENSOR WITH Stab-ElTOPTION: Neutralizes trace contaminants including acids such as CO2, H2S, NOX, SOX, etc. (Consult Delta F for concentration limits.)

Options
  • Oxygen Alarms
  • Stab-ElTAcid Gas System
  • Remote Operation
  • Remote Sensor
  • Flow Alarm
  • Three Range
  • Sample Pump

Accuracy (at constant temperature)
  • Greater of ±3% of reading or 0.5% of range

Response Time
  • Responds instantaneously to Oxygen change. Equilibrium depends on specific conditions. Typically <10 seconds to read 90% of a step change.

Sample Requirements
  • Gas phase, non-condensing, 0° to 150° F.
  • Flow should be 1.0 to 3.0 scfh at 0.2 to 1.0 psig pressure.

Misc.
  • Operate between 0° and 45°C ambient (32° and 120° F)
  • 115 VAC or 220 VAC at 25 Watts