sigma logo

DF-150E Custom Configure the Exact Oxygen Analyzer You Need




The DF-150 features:
  • Maximum configuration flexibility - build it the way you need it
  • Superior performance out-of-the-box
  • Accurate - ±2% full scale at constant temperature
  • Instantaneous response to oxygen change
  • Fast response - typically less than 10 seconds to read 90% of a step change

Every DF-150E is custom configured under ISO-9001 controls to your precise specifications and calibrated to standards traceable to NIST and operated for several days to ensure flawless performance. The ultra-stable E-Sensor can operate for years without re-calibration giving you a long and useful life.

DF-150E Specifications Range
  • Ranges are available from 0-10 ppm to 0-25%

Sample Gas Compatibility
  • STANDARD SENSOR: All inert and passive gases including N2, CO, H2, Ar, freons, hydrocarbons, etc.
  • SENSOR WITH Stab-ElT OPTION: Neutralizes trace contaminants including acids such as CO2, H2S, NOX, SOX, etc. (Consult Delta F for concentration limits.)

Options
  • Oxygen Alarms
  • Stab-ElT Acid Gas System
  • CE Certification
  • Analog and Digital Outputs
  • Classified Area Enclosures
  • Flow Alarm
  • Three range
  • Sample Pump
  • Remote Sensor

Accuracy (at constant temperature)
  • Greater of ±3% of reading or 0.5% of range

Response Time
  • Responds instantaneously to O2 change. Equilibrium depends on specific conditions. Typically <10 seconds to read 90% of a step change.

Sample Requirements
  • Gas phase, non-condensing, 0° to 150° F.
  • Flow should be 1.0 to 3.0 scfh at 0.2 to 1.0 psig pressure.

Misc.
  • 19.7cm W x 21.6cm H x 25.4cm D 7kg. (7.75"W x 8.5"H x 10"D 15 lbs.)
  • Operate between 0° and 45° C ambient (32° and 120° F)
  • 115 VAC or 220 VAC at 25 Watts

For more information, click here to view the DF-150E datasheet pdf.